Multicenter Validation of a Urine CXCL10 Assay for Noninvasive Monitoring of Renal Transplants

期刊

TRANSPLANTATION
卷 107, 期 7, 页码 1630-1641

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/TP.0000000000004554

关键词

-

向作者/读者索取更多资源

This study demonstrates that urine CXCL10 Luminex immunoassay is robust, reproducible, and accurate, indicating its potential for clinical use in HLA laboratories for posttransplant surveillance.
Background. Urine CXCL10 (C-X-C motif chemokine ligand 10, interferon gamma-induced protein 10 [IP10]) outperforms standard-of-care monitoring for detecting subclinical and early clinical T-cell-mediated rejection (TCMR) and may advance TCMR therapy development through biomarker-enriched trials. The goal was to perform an international multicenter validation of a CXCL10 bead-based immunoassay (Luminex) for transplant surveillance and compare with an electrochemiluminescence-based (Meso Scale Discovery [MSD]) assay used in transplant trials. Methods. Four laboratories participated in the Luminex assay development and evaluation. Urine CXCL10 was measured by Luminex and MSD in 2 independent adult kidney transplant trial cohorts (Basel and TMCT04). In an independent test and validation set, a linear mixed-effects model to predict (log(10)-transformed) MSD CXCL10 from Luminex CXCL10 was developed to determine the conversion between assays. Net reclassification was determined after mathematical conversion. Results. The Luminex assay was precise, with an intra- and interassay coefficient of variation 8.1% and 9.3%; showed modest agreement between 4 laboratories (R 0.96 to 0.99, P < 0.001); and correlated with known CXCL10 in a single- (n = 100 urines, R 0.94 to 0.98, P < 0.001) and multicenter cohort (n = 468 urines, R 0.92, P < 0.001) but the 2 assays were not equivalent by Passing-Bablok regression. Linear mixed-effects modeling demonstrated an intercept of -0.490 and coefficient of 1.028, showing Luminex CXCL10 are slightly higher than MSD CXCL10, but the agreement is close to 1.0. After conversion of the biopsy thresholds, the decision to biopsy would be changed for only 6% (5/85) patients showing acceptable reclassification. Conclusions. These data demonstrate this urine CXCL10 Luminex immunoassay is robust, reproducible, and accurate, indicating it can be readily translated into clinical HLA laboratories for serial posttransplant surveillance.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据