All-in-one microfluidic device with an integrated porous filtration membrane for on-site detection of multiple salivary biomarkers

期刊

SENSORS AND ACTUATORS B-CHEMICAL
卷 379, 期 -, 页码 -

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2022.133214

关键词

Point-of-care sensor; Microfluidic sensor; Salivary biomarkers; 3D-printing; Screen-printing; Wireless data collection

向作者/读者索取更多资源

A hand-held microfluidic device with integrated filtration membrane is developed for electrochemical monitoring of multiple salivary biomarkers. The device uses a "channel-inside-channel" configuration to enable pump-less passive flow of saliva. It also integrates a flexible porous filtration membrane at the inlet to prevent biofouling. Enzyme-based electrochemical sensors are used to monitor glucose, lactate, cholesterol, and uric acid levels in saliva samples. The device shows good correlation with blood glucose measurements, demonstrating its efficacy in saliva-based point-of-care analysis.
A hand-held microfluidic device is developed with an integrated filtration membrane for electrochemical monitoring of multiple salivary biomarkers. The platform consists of array of screen-printed electrodes, inte-grated with a high throughput microfluidic system that was developed via clean-room free 3D-printing tech-nique. A 'channel-inside-channel' configuration is introduced to facilitate pump-less passive flow of saliva at the broad sensing chambers. In order to eliminate the salivary protein based biofouling problems, a flexible porous filtration membrane is integrated at the microfluidic inlet that avoids the passage of bulkier protein constituents. As the presence of separate insulator layer on the sensing electrodes adversely affects its integration with the fluidic channel, the electrode substrate itself is utilized as the insulator by exposing the bottom region of the printed electrodes as active region. Enzyme based electrochemical sensors are developed with suitable nano -materials for monitoring glucose, lactate, cholesterol and uric acid from collected saliva samples. The device showed distinct differences in salivary glucose levels for healthy subjects before and after consuming glucose-rich drinks. The glucose sensor responses are validated with high performance liquid chromatography and com-mercial colorimetric assay methods, and observed good correlation with parallel finger prick blood glucose measurements with an average Pearson coefficient of 0.9161, demonstrating its efficacy in saliva based point-of -care analysis.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据