4.4 Article

Performance of Roche CAP/CTM HIV-1 qualitative test version 2.0 using dried blood spots for early infant diagnosis

期刊

JOURNAL OF VIROLOGICAL METHODS
卷 229, 期 -, 页码 12-15

出版社

ELSEVIER
DOI: 10.1016/j.jviromet.2015.12.006

关键词

HIV-1; Early infant diagnosis; CAP/CTM v 2.0; DBS

向作者/读者索取更多资源

In the context of early infant diagnosis (EID) decentralization in sub-Saharan Africa, dried blood spot (DBS) is now widely used for HIV proviral DNA detection in resource-limited settings. A new version of CAP/CTM (version 2) has been introduced, recently by Roche Diagnosis as a new real-time PCR assay to replace previous technologies on qualitative detection of HIV-1 DNA using whole blood and DBS samples. The objective of this study was to evaluate CAP/CTM version 2 compared to CAP/CTM version 1 and Amplicor on DBS. A total of 261 DBS were collected from children aged 4 weeks to 17 months born from HIV-seropositive mothers and tested by the three techniques. CAP/CTM version 2 showed 100% of agreement with Amplicor including 74 positive results and 187 negative results. CAP/CTM version 2 versus CAP/CTM version 1 as well as CAP/CTM version 1 versus Amplicor showed two discordant results giving a sensitivity of 98.6%, specificity of 99.5%, positive predictive value of 98.6% and negative predictive value of 99.5%. The concordance was 99.12% (95% of confidence interval) giving a Kappa coefficient of 0.97 (p < 0.001). These findings confirmed the expected good performance of CAP/CTM version 2 for HIV-1 EID. (C) 2015 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据