4.6 Article

Patient-specific Monte Carlo-based organ dose estimates in spiral CT via optical 3D body scanning and adaptation of a voxelized phantom dataset: proof-of-principle

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 68, 期 8, 页码 -

出版社

IOP Publishing Ltd
DOI: 10.1088/1361-6560/acc37d

关键词

CT dosimetry; voxelized phantoms; Monte Carlo simulations; organ doses; Geant4; 3D scanner

向作者/读者索取更多资源

This study presents a method for personalized organ dose estimates before CT exams using 3D optical body scanning and Monte Carlo simulations. By adapting a reference phantom to the body size and shape measured with a portable 3D optical scanner, a voxelized phantom was derived. The internal body anatomy derived from a phantom dataset matched for gender, age, weight, and height was incorporated into the voxelized body phantom to provide more accurate organ dose estimates.
Objective. We present a method for personalized organ dose estimates obtained before the computed tomography (CT) exam, via 3D optical body scanning and Monte Carlo (MC) simulations. Approach. A voxelized phantom is derived by adapting a reference phantom to the body size and shape measured with a portable 3D optical scanner, which returns the 3D silhouette of the patient. This was used as an external rigid envelope for incorporating a tailored version of the internal body anatomy derived from a phantom dataset (National Cancer Institute, NIH, USA) matched for gender, age, weight, and height. The proof-of-principle was conducted on adult head phantoms. The Geant4 MC code provided estimates of the organ doses from 3D absorbed dose maps in the voxelized body phantom. Main results. We applied this approach for head CT scanning using an anthropomorphic voxelized head phantom derived from 3D optical scans of manikins. We compared the estimates of head organ doses with those provided by the NCICT 3.0 software (NCI, NIH, USA). Head organ doses differed up to 38% using the proposed personalized estimate and MC code, with respect to corresponding estimates calculated for the standard (non-personalized) reference head phantom. Preliminary application of the MC code to chest CT scans is shown. Real-time pre-exam personalized CT dosimetry is envisaged with adoption of a Graphics Processing Unit-based fast MC code. Significance. The developed procedure for personalized organ dose estimates before the CT exam, introduces a new approach for realistic description of size and shape of patients via voxelized phantoms specific for each patient.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据