Simple, non-mechanical and automatic calibration approach for axial-scanning microscopy with an electrically tunable lens

期刊

MICROSCOPY RESEARCH AND TECHNIQUE
卷 86, 期 10, 页码 1391-1400

出版社

WILEY
DOI: 10.1002/jemt.24337

关键词

autofocusing; calibration; microscopy; non-mechanical; tunable lenses

向作者/读者索取更多资源

We present a simple and robust calibration approach for axial-scanning microscopy using an electrically tunable lens (ETL). By capturing multiple images at different ETL currents and evaluating their sharpness, we calibrate the dependence of the axial focus shift on the ETL current. Additionally, we measure the period of repeating lines in the captured images to calibrate the dependence of microscope magnification on the ETL current. This approach enables simultaneous calibration of axial focus shift and magnification, without the need for mechanical scanning.
We describe a simple and robust calibration approach for axial-scanning microscopy that realizes axial focus shifts with an electrically tunable lens (ETL). We demonstrate the calibration approach based on a microscope with an ETL placed close to the rear stop of the objective lens. By introducing a target-consisted of repeating lines at one known frequency and placed at a similar to 45 degrees angle to the imaging path, the calibration method captures multiple images at different ETL currents and calibrates the dependence of the axial focus shift on the ETL current by evaluating the sharpness of the captured images. It calibrates the dependence of the magnification of the microscope on the ETL current by measuring the period of the repeating lines in the captured images. The experimental results show that different from the conventional calibration procedure, the proposed scheme does not involve any mechanical scanning and can simultaneously calibrate the dependence of the axial focus shift and the magnification on the ETL current. This might facilitate imaging studies that require the measurement of fine structures in a 3D volume. We also show the calibration procedure can be used to estimate the radius of a conner-arc sample, fabricated using laser micromachining. We believe that this easy-to-use calibration approach may facilitate use of ETLs for a variety of imaging platforms. It may also provide new insights for the development of novel 3D surface measurement methods.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据