4.6 Article

A novel MR-conditional cam-based automatic prostate biopsy device

期刊

MECHATRONICS
卷 91, 期 -, 页码 -

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.mechatronics.2023.102966

关键词

MR conditional; Prostate biopsy; Medical robotics

向作者/读者索取更多资源

This paper presents an automatic biopsy device that can be used in the MRI environment and works with a MR conditional robotic prostate biopsy system. The device, based on a compact cam design, can perform all actions required for prostate biopsy using a single actuator. Experimental evaluation showed that the device can easily and completely perform tissue sampling, with comparable performance to existing manual biopsy guns. MRI compatibility testing demonstrated a 18.7% reduction in signal-to-noise ratio during device operation.
Biopsy is the most commonly used method for diagnosing prostate cancer, in which a biopsy gun is used to obtain the tissue samples. The outstanding performance of magnetic resonance imaging (MRI) for soft tissue can provide better medical guidance for prostate biopsy. This paper reports an automatic biopsy device that can be used in the MRI environment and that is expected to be used together with the MR conditional robotic prostate biopsy system. This paper proposes a compact cam-based automatic biopsy device that can perform all actions required for prostate biopsy sampling procedure using a single actuator. Its design concept, implementation, required force analysis, sampling performance evaluation, and MR compatibility evaluation are also provided. The experimental evaluation showed that the tissue sampling process can be easily and completely performed using the proposed automatic biopsy device, and that its sampling performance is comparable to that of the existing commercially available manual biopsy gun. MR compatibility test showed that an 18.7% reduction in the signal-to-noise ratio when the device was operating.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据