4.5 Article

Live-view 4D GRASP MRI: A framework for robust real-time respiratory motion tracking with a sub-second imaging latency

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 90, 期 3, 页码 1053-1068

出版社

WILEY
DOI: 10.1002/mrm.29700

关键词

GRASP; live-view; MRI-Linac; navi-stack-of-stars; real-time imaging; respiratory motion

向作者/读者索取更多资源

This study proposes a framework called live-view golden-angle radial sparse parallel (GRASP) MRI for low-latency and high-fidelity real-time volumetric MRI. The framework places the burden of acquisition and reconstruction in the off-view stage, enabling low-latency real-time 3D imaging in the live-view stage. The accuracy of live-view GRASP MRI and the robustness of 2D navigators for characterizing respiratory variations and/or body movements were assessed.
Purpose: To propose a framework called live-view golden-angle radial sparse parallel (GRASP) MRI for low-latency and high-fidelity real-time volumetric MRI.Methods: Live-view GRASP MRI has two stages. The first one is called an off-view stage and the second one is called a live-view stage. In the off-view stage, 3D k-space data and 2D navigators are acquired alternatively using a new navi-stack-of-stars sampling scheme. A 4D motion database is then generated that contains time-resolved MR images at a sub-second temporal resolution, and each image is linked to a 2D navigator. In the live-view stage, only 2D navigators are acquired. At each time point, a live-view 2D navigator is matched to all the off-view 2D navigators. A 3D image that is linked to the best-matched off-view 2D navigator is then selected for this time point. This framework places the typical acquisition and reconstruction burden of MRI in the off-view stage, enabling low-latency real-time 3D imaging in the live-view stage. The accuracy of live-view GRASP MRI and the robustness of 2D navigators for characterizing respiratory variations and/or body movements were assessed.Results: Live-view GRASP MRI can efficiently generate real-time volumetric images that match well with the ground-truth references, with an imaging latency below 500 ms. Compared to 1D navigators, 2D navigators enable more reliable characterization of respiratory variations and/or body movements that may occur throughout the two imaging stages.Conclusion: Live-view GRASP MRI represents a novel, accurate, and robust framework for real-time volumetric imaging, which can potentially be applied for motion adaptive radiotherapy on MRI-Linac.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据