Achieving Sub-Parts-per-Million Mass Measurement Accuracy on an Orbitrap Mass Spectrometry Imaging Platform without Automatic Gain Control

期刊

出版社

AMER CHEMICAL SOC
DOI: 10.1021/jasms.3c00004

关键词

centroid; mass measurement accuracy; IR-MALDESI; mass spectrometry imaging; Orbitrap; automatic gain control

向作者/读者索取更多资源

Profile data collection in mass spectrometry imaging (MSI) studies often results in large files and long processing times. This study compares profile and centroid data collection methods in high-resolution accurate mass MSI. The results show that centroid data collection improves mass measurement accuracy and lipid annotation sensitivity compared to profile data collection.
The collection of profile data is standard practice within the field of mass spectrometry (MS). However, profile data collection often results in large data files that require extensive processing times, especially in mass spectrometry imaging (MSI) studies where thousands of high-resolution scans are recorded. Natively collecting centroid MS data is an alternative that effectively reduces both the resulting file size and the data processing time. Herein, high-resolution accurate mass (HRAM) Orbitrap MSI data on mouse liver tissue sections without automatic gain control (AGC) were natively collected in both profile and centroid modes and compared based on the file size and processing time. Additionally, centroid data were evaluated against the profile data with regard to the spectra integrity, mass measurement accuracy (MMA), and the number of lipid annotations to ensure that centroid data did not compromise the data quality. For both native and postacquisition centroided data, the variation in mass measurement accuracy decreased relative to the profile data collection. Furthermore, centroid data collection increased the number of METASPACE database annotations indicating higher sensitivity and greater accuracy for lipid annotation compared to native profile data collection. Profile MSI data was shown to have a higher likelihood of false positive identifications due to an increased number of data points on either side of the peaks, whereas the same trend was not observed in data collected in native centroid data collection. This publication explores and explains the importance in properly centroiding MSI data, either natively or by adequate centroiding methods, to obtain the most accurate information and come to the best conclusions. These data support that natively collecting centroid data significantly improves MMA to sub-ppm levels without AGC and reduces false positive annotations.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据