4.7 Article

Parameter Estimation of the Single-Dispersion Fractional Cole-Impedance Model With the Embedded Hardware

期刊

IEEE SENSORS JOURNAL
卷 23, 期 12, 页码 12978-12987

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/JSEN.2023.3269952

关键词

Bioimpedance; Cole equation; equivalent circuits; estimation; fractional-order circuits

向作者/读者索取更多资源

This article presents a low-complexity method for parameter estimation of the Cole-impedance model suitable for low-cost embedded hardware. The proposed method uses only the measured real and imaginary impedance, without any specific software package/toolbox, or initial values provided by the user. The method has been validated and compared with other relevant work and commercial software, confirming its reliability for bioimpedance processing on embedded hardware.
Bioimpedance modeling with equivalent electrical circuits has an important role in various biomedical applications, as it facilitates understanding of underlying physical and electrochemical processes in applications such as body composition measurements and assessment of clinical conditions. However, the estimation of model parameter values is not a straightforward task, especially when complex circuits with fractional-order components [e.g., constant phase elements (CPEs)] are used. In this article, we propose a low-complexity method for parameter estimation of the Cole-impedance model suitable for low-cost embedded hardware (e.g., 8-bit microcontrollers). Our approach uses only the measured real and imaginary impedance, without any specific software package/toolbox, or initial values provided by the user. The proposed method was validated with synthetic (noiseless and noisy) data and experimental right-side, hand-to-foot bioimpedance data from a healthy adult participant. Moreover, the proposed method was compared in terms of accuracy with the recently published relevant work and commercial Electrical Impedance Spectroscopy software (Bioimp 2.3.4). The performance evaluation in terms of complexity (suitable for deployment for the microcontroller-based platform with 256 kB of RAM and 16 MHz clock speed), execution time (18 s for the dataset with 256 points), and cost (<$25) confirms the proposed method in regards to reliable bioimpedance processing using embedded hardware.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据