Fundus imaging and perimetry in patients with idiopathic intracranial hypertension-an intermethod and interrater validity study

期刊

EUROPEAN JOURNAL OF NEUROLOGY
卷 30, 期 7, 页码 1973-1982

出版社

WILEY
DOI: 10.1111/ene.15802

关键词

fundus images; idiopathic intracranial hypertension; intermethod assessment; intrarater assessment; papilledema; perimetry; validation study; visual field assessment

向作者/读者索取更多资源

This study validated the COMPASS system, a fundus imaging and perimetric visual field assessment system, for diagnosing papilledema in patients suspected of idiopathic intracranial hypertension. The system showed a sensitivity of 87% and specificity of 73% in detecting papilledema. However, it had a lower sensitivity of 59% and moderate agreement in detecting visual field defects compared to the assessments made by neuroophthalmologists.
Background and purpose: There is a need to improve the diagnostic process of patients suspected of papilledema. In patients with known or suspected idiopathic intracranial hypertension a fundus imaging and perimetric visual field assessment system (COMPASS) performed at a headache center was validated in comparison to an assessment (Topcon plus OCTOPUS) at a neuroophthalmological clinic.Methods: For intermethod assessment, blinded fundus images and perimetry from COMPASS versus Topcon plus OCTOPUS were assessed by a neuroophthalmologist. For interrater assessment, fundus images and perimetry obtained by the COMPASS system were assessed by an untrained medical doctor, a trained neurologist and a trained medical student and compared to the neuroophthalmologist's assessments.Results: For the intermethod variation of the presence of papilledema on fundus images, a kappa value of 0.60, sensitivity of 87% and specificity of 73% were found. The interrater variation of the presence of papilledema on fundus images showed kappa values ranging from 0.43 to 0.74, sensitivity values ranging from 70% to 96% and specificity values ranging from 46% to 93% when comparing the assessments made by the headache center staff with neuroophthalmologist's assessments. The COMPASS showed a 59% sensitivity and moderate agreement in detecting visual field defects compared with OCTOPUS. The visual field assessment showed only slight to fair agreement from 0.19 to 0.31 between assessments made by the headache center staff and the neuroophthalmologist.Conclusion: The COMPASS system can be used with reasonable sensitivity in the assessment of papilledema in patients suspected of idiopathic intracranial hypertension at a tertiary headache center.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据