Accuracy of immediate dental implant placement with task-autonomous robotic system and navigation system: An in vitro study

期刊

CLINICAL ORAL IMPLANTS RESEARCH
卷 -, 期 -, 页码 -

出版社

WILEY
DOI: 10.1111/clr.14104

关键词

computer-assisted; dental implants; immediate implant placement; surgery

向作者/读者索取更多资源

This study aimed to compare the accuracy of dental implant placement using a task-autonomous robotic system and a dynamic navigation system. The results showed that the robotic system had more precise placement in both immediate and conventional implant procedures. However, further clinical trials are needed to confirm its performance.
Objectives: The aim of this study was to compare the accuracy of dental implant placement in a single tooth gap, including the postextraction site and healed site, using a task-autonomous robotic system and a dynamic navigation system. Materials and Methods: Forty partially edentulous models requiring both immediate and conventional implant placement were randomly divided into a robotic system group and a navigation system group. The coronal, apical, and angular deviations of the implants were measured and assessed between the groups. Results: The deviations in immediate implant placement were compared between the robotic system and dynamic navigation system groups, showing a mean (+/- SD) coronal deviation of 0.86 +/- 0.36 versus 0.70 +/- 0.21 mm (p = .101), a mean apical deviation of 0.77 +/- 0.34 versus 0.95 +/- 0.38 mm (p = .127), and a mean angular deviation of 1.94 +/- 0.66 degrees versus 3.44 +/- 1.38 degrees (p < .001). At the healed site, significantly smaller coronal deviation (0.46 +/- 0.29 vs. 0.70 +/- 0.30 mm, p = .005), apical deviation (0.56 +/- 0.30 vs. 0.85 +/- 0.25 mm, p < .001), and angular deviation (1.36 +/- 0.54 vs. 1.80 +/- 0.70 mm, p = .034) were found in the robotic system group than in the dynamic navigation group. Conclusions: The position in both immediate and conventional implant placement was more precise with the task-autonomous robotic system than with the dynamic navigation system. Its performance in actual clinical applications should be confirmed in further trials.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据