4.5 Article

Performance and reliability of two frequently used point-of-care blood gas analyzers at 423 and 4,559 m

期刊

CLINICAL BIOCHEMISTRY
卷 116, 期 -, 页码 38-41

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.clinbiochem.2023.03.007

关键词

Hypoxia; High altitude; Carboxyhemoglobin; Carbon monoxide rebreathing

向作者/读者索取更多资源

This study investigated the reliability and robustness of two widely used blood gas analyzers at high altitude. The results showed that compared to RAD, SIE had lower reliability at both 423m and 4559m, especially at 4559m where data reporting issues were more prevalent.
Background and Objectives: Blood gas analyzers (BGA) aid medical decision-making. Their specified performance criteria are based on sea level conditions. However, millions of people are living at high altitude (HA) where the performance of BGAs is poorly characterized. We investigated the effect of exposure to 4,559 m on the reliability and robustness of two BGAs widely used at HA. Methods: In this prospective study arterial blood samples from 13 volunteers (2 female) with susceptibility to the development of high-altitude pulmonary edema were collected once near sea level at 423 m (nSL423) and three times at high altitude (HA4,559). Samples were measured in triplicate with the cartridge BGAs Rapidpoint 500 (SIE; Siemens Healthcare) and the ABL90 (RAD; Radiometer) to calculate coefficients of variation (CV) and intraclass correlation coefficients (ICC) within a mixed model. Results: At nSL423 and HA4,559, 3% and 17% of all data were not reported with SIE, mainly due to clotting of the sample caused by delays because of the frequent automated calibration routines. No data were missing with RAD. ICCs were not significantly lower (mean (min-max) 0.87 (0.68-0.98) vs. 0.94 (0.84-1.00); p = 0.217) with SIE at nSL423, but significantly lower at HA4,559 (0.87 (0.49-1.00) vs. 0.99 (0.96-1.00); p = 0.025). All CVs, except that for arterial oxygen saturation at HA4,559,were higher with SIE . Conclusion: In this study, the reliability of RAD was superior to SIE at nSL423 and HA4,559. In contrast to RAD, the performance of SIE declined at HA4,559. SIE was more prone to not reporting all variables, especially at HA4559.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据