Analytical Performance Evaluation for Estradiol using Liquid Chromatography-Tandem Mass Spectrometry

期刊

CLINICAL BIOCHEMISTRY
卷 113, 期 -, 页码 59-63

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.clinbiochem.2023.01.003

关键词

Estradiol; Ultra -performance liquid chromatography; Tandem mass spectrometry

向作者/读者索取更多资源

In this study, a highly sensitive and derivatization-free LC-MS/MS method was successfully developed for measuring estradiol levels in patients with estradiol-related endocrine disorders. The method showed good analytical performance and comparable results to a chemiluminescent microparticle immunoassay.
Background: There is an increased need for the sensitive and accurate measurement of estradiol levels in patients with estradiol-related endocrine disorders. Liquid chromatography-tandem mass spectrometry (LC-MS/MS) is a reliable and accurate method for measuring steroid hormone levels. Here, we aimed to establish an LC-MS/MSbased method to quantify estradiol levels without sample derivatization and evaluated its analytical performance.Methods: Sciex Triple Quad 6500 + LC-MS/MS was used for estradiol analysis. We evaluated its analytical performance, including linearity, precision, the lower limit of detection and quantification (LoD and LoQ, respectively), accuracy, and carryover. The estradiol results determined by LC-MS/MS were compared with those obtained using a chemiluminescent microparticle immunoassay.Results: The LC-MS/MS output was linear for serum estradiol concentrations in the range of 0.2-10311.6 pmol/L. The intra-laboratory precision (coefficient of variation) was 3.0-10.1 %. The LoD and LoQ were 2.8 and 7.5 pmol/L, respectively. The overall accuracy was within 15 % of bias, and the carryover was within the acceptable range (<1.0 %). The results of the estradiol analysis determined by LC-MS/MS were comparable to those obtained by the chemiluminescent microparticle immunoassay (r(2) = 0.9843), although there was a negative bias of - 17.82 (95 % confidence interval, -27.21 to - 8.44).Conclusions: A highly sensitive, derivatization-free LC-MS/MS method was successfully developed in this study. This may be beneficial for estradiol measurements in patients with estradiol-related endocrine disorders.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据