Non-linearity in lipase assays: A multicentric comparison on different analysers

期刊

CLINICAL BIOCHEMISTRY
卷 114, 期 -, 页码 63-66

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.clinbiochem.2023.02.003

关键词

Lipase; Linearity; Analysers; Analytical; Assay method

向作者/读者索取更多资源

This study aims to investigate the non-linearity and distribution differences in lipase assays on different analysers. By evaluating 18 plasma samples on various analysers, significant non-linearity biases were observed in the Roche Cobas c503 method and the Randox method on the Abbott Alinity.
Objectives: Non-linearity in lipase assays and the ensuing gaps in results distribution have been described on Roche analysers, but have yet to be studied on other analysers.Design and methods: Eighteen lithium-heparinized plasma pools of lipase activities decreasing from 1700 to <4 U/ L were prepared for multicentric evaluation on several analysers. Non-linearity was modelled as the difference between the polynomial regression of lipase activities depending on relative dilutions over the primary measuring range, and the linear regression of the same variables above the manufacturer's limit of linearity (MLL). Gaps in lipase distribution resulting from non-linearity were graphically evidenced through histograms. Upper limits of gaps were calculated, which are lipase activities where non-linearity biases no longer impact the diluted lipase results.Results: MLLs and lipase (U/L) calculated at MLL (%biases versus MLL) were respectively: 1200 and 1124 (-6.3%) on the Architect C16000 (Abbott); 300 and 248 (-17.3%) on the Cobas c503 (Roche); 1500 and 1458 (-2.8%) on the Dimension Vista (Siemens); and 700 and 659 (-5.9%) on the Atellica CH930 (Siemens). Using Sentinel Lipase reagents on Abbott analysers, these measurements were respectively: 300 and 294 (-2.0%) on the Architect C16000, and 300 and 298 (-0.7%) on the Alinity. Setting Randox Lipase reagents on the Alinity, MLL and lipase at MLL were 953 and 776 (-18.6%), respectively.Conclusions: Considering the desirable (+/- 14.2 %) and optimal (+/- 7.1 %) allowable total error for lipase (EFLM/ EuBIVAS), biases at manufacturer's limit of linearity were acceptable, except for Roche Cobas c503 method and Randox method on Abbott Alinity.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据