Filter Fluorometer Calibration Without the Fluorometer

期刊

APPLIED SPECTROSCOPY
卷 77, 期 9, 页码 1053-1063

出版社

SAGE PUBLICATIONS INC
DOI: 10.1177/00037028231181593

关键词

Fluorescence; chlorophyll; fluorometer; calibration; Raman; water; polarization; instrumentation; modeling; aberrations

向作者/读者索取更多资源

We have developed a lightweight, low-power, waterproof filter fluorometer for chlorophyll-a detection. Through modeling and calibration, we determined the calibration slope and detection limit for dissolved chl-a in water. The experimental results validate our model, showing good agreement with the estimated values. The achieved detection limit is suitable for practical applications.
We recently described a lightweight, low-power, waterproof filter fluorometer using a 180 & DEG; backscatter geometry for chlorophyll-a (chl-a) detection. Before it was constructed it was modeled to ensure it would have satisfactory performance. This manuscript repeats the modeling process that allows the calibration slope and detection limit for a fluorescent analyte in water to be estimated from system component performance and conventional spectrofluorometry alone. These values are validated by comparison to the experimental result of calibration from the completed instrument. Our model yields a calibration slope of 8.22 mV-L/& mu;g for dissolved chl-a, consistent with the experimentally measured slope of 8.21 mV-L/& mu;g. The detection limit modeled from this slope and an estimate of the baseline noise of the instrument was 0.15 & mu;g/L chl-a, while the measured detection limit using real blank samples was 0.18 & mu;g/L, in 0.1 s differential measurements.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据