4.5 Article

Thermal injury of the recurrent laryngeal nerve by THUNDERBEAT during thyroid surgery: findings from continuous intraoperative neuromonitoring in a porcine model

期刊

JOURNAL OF SURGICAL RESEARCH
卷 200, 期 1, 页码 177-182

出版社

ACADEMIC PRESS INC ELSEVIER SCIENCE
DOI: 10.1016/j.jss.2015.06.066

关键词

Thyroid; Recurrent laryngeal nerve; Continuous neuromonitoring; Energy-based device

类别

资金

  1. Technology Innovation Program (Industrial Strategic Technology Development Program) [10049743]
  2. Ministry of Trade, Industry and Energy (MI, Korea)

向作者/读者索取更多资源

Background: Recurrent laryngeal nerve (RLN) palsy is the most common and serious complication of thyroid surgery. The use of energy-based devices (EBDs) has replaced hand-tying methods in many institutions. However, EBD use proximal to the RLN presents risks related to lateral thermal spread and associated nerve damage. THUNDERBEAT (TB) is one of the most widely used EBDs. This study aimed to test the safety of TB during thyroidectomy. Methods: Four piglets weighing 30-40 kg experienced thyroidectomy while continuous electrophysiologic monitoring (continuous intraoperative neuromonitoring) occurred, using an electromyography endotracheal tube and NIM 3.0 response system. TB was applied at various distances from the RLN, and we assessed the safety of the protocols. Results: Adverse electromyography events did not occur at distances >3 mm from the RLN. Amplitude decreased at 2 mm from the RLN after 8 s. However, immediate loss of signal occurred at 1 mm from the RLN, likely due to immediate shrinkage of surrounding tissue after TB application. Conclusions: TB can be used safely at 3 mm from the RLN but must be used for <8 s at more proximal locations. This is the first report assessing the safety of TB, and findings indicate that TB should be used at least 1 mm from the RLN to avoid injury. (C) 2016 Elsevier Inc. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据