4.7 Article

Validation of a fully automated chemiluminescent immunoassay for cattle serum and plasma progesterone measurement

期刊

FRONTIERS IN VETERINARY SCIENCE
卷 9, 期 -, 页码 -

出版社

FRONTIERS MEDIA SA
DOI: 10.3389/fvets.2022.1064201

关键词

cattle; progesterone; immunoassay; validation; serum; plasma

向作者/读者索取更多资源

The automated IPI is a precise and accurate method for measuring progesterone levels in both serum and plasma of cattle. It exhibits linearity and high precision, making it suitable for serum and plasma samples.
Introduction: Monitoring circulating progesterone concentration ([P4]) is an important component of basic and applied reproduction research and clinical settings. IMMULITE (R) 2000 XPi (Siemens Healthineers, Cary, NC) is a newly upgraded fully automated immunoassay system marketed for human use to measure concentrations of different measurands including P4. Objectives: Our objective was therefore to characterize the analytical performance of the IMMULITE (R) 2000 XPi P4 immunoassay (IPI) across the reportable range in serum and plasma of cattle. Methods: The IPI validation protocols included characterization of the method linearity, within-run, and between-run precision through calculation of the coefficient of variation (CV). The method accuracy was assessed through the calculation of the spiking-recovery (SR) bias across the reportable range (0.2-40.0 ng/mL). Passing-Bablok regression and Bland-Altman plots were used to determine the interlaboratory bias for two laboratories. Three types of observed total error (TEo) were calculated based on the considered type of bias, TEo(SR) (spiking-recovery), TEo(RB) (range-based bias), and TEo(AB) (average-based bias). Results: The IPI was linearly related to the true value (R-2 = 0.997) across the reportable range. The within-run and between-run precision (CV%) of the IPI for both serum and plasma [P4] of clinical relevance (1, 2, 5, and 10 ng/mL) were <5 and <10%, respectively. The TEo reported here for serum and plasma at [P4] of 1 and 5 ng/mL was similar to 20 and 25%, respectively. Of interest, the three types of TEo were relatively similar regardless of the considered bias. Conclusions: We concluded that the automated IPI provides a precise, accurate, reliable, and safe method for measuring [P4] in both serum and plasma of cattle. Consistent with the manufacturer's recommendations, the serum matrix is more accurate than plasma.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据