4.3 Article

OPUS: an easy way to push the limits of SANS instruments towards USANS

期刊

JOURNAL OF APPLIED CRYSTALLOGRAPHY
卷 55, 期 -, 页码 1640-1651

出版社

INT UNION CRYSTALLOGRAPHY
DOI: 10.1107/S1600576722008834

关键词

ultra-small-angle neutron scattering (USANS); pin-hole small-angle neutron scattering (SANS); neutron ray tracing; McStas simulations; neutron lenses; LoKI; D11

向作者/读者索取更多资源

Chemistry and physics have made significant advances in recent years, leading to the development of complex systems. To understand the structure of these new materials, specific characterization tools are required. Simultaneous small-angle neutron scattering (SANS) and ultra-small-angle neutron scattering (USANS) measurements provide a unique approach to study this complexity. This paper presents the design of a USANS option for SANS instruments, which allows for the study of different complex systems under the same experimental conditions. Simulations show that the multi-beam approach can extend the Q range to lower values.
Chemistry and physics have made major advances in recent years, yielding much more complex systems with high hierarchical order across multiple length scales. Accordingly, characterization tools are required that can elucidate the structure of such new materials over all length scales. Simultaneous small-angle neutron scattering (SANS) and ultra-small-angle neutron scattering (USANS) measurements are a unique tool to study such complexity and can be applied to very different fields of science. The OPUS (Option USANS) project is the study of a USANS option for SANS instruments, designed to be very versatile and easy to implement. The main idea is to provide the opportunity to study at the same time, and under the same experimental conditions, complex systems such as polymers, bio-systems, complex fibres and self-assembling systems. More specifically, this work presents the design of an option that could be applied to the suite of SANS instruments at the European Spallation Source (ESS) which will allow exploration of a Q range with a minimum Q down to one order of magnitude lower than the value attainable with the standard SANS instrument at the ESS. The proposed setup, based on the SAMBA (smallangle multi-beam analysis) approach, is very easy and fast to implement on a conventional SANS instrument and constitutes a multi-beam approach involving two multi-slits and a set of lenses near the sample position. This contribution describes all the focusing elements necessary to attain the proposed configuration and a detailed study using McStas simulations to optimize all the parameters involved for two SANS instruments: the future LoKI at the ESS and the present D11 at the Institut Laue-Langevin, the latter used as a benchmark for the model. Simulations performed without taking into account gravity effects show that the multi-beam approach allows extending the Q ranges to 9 x 10(-5)-7 x 10(-4) angstrom(-1) and 5 x 10(-5)-3 x 10(-4) angstrom(-1) for LoKI and D11, respectively.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据