Inter-rater and intra-rater reliability of isotonic exercise monitoring device for measuring active knee extension

期刊

PEERJ
卷 11, 期 -, 页码 -

出版社

PEERJ INC
DOI: 10.7717/peerj.14672

关键词

Isotonic exercise device; Knee extension; Inter-rater reliability; Intra-rater reliability; Knee rehabilitation; Surface electromyography; Range of motion

向作者/读者索取更多资源

This study assessed the reliability of electromyography and range of motion measurements obtained using a knee exercise monitoring system. The results showed moderate to excellent inter- and intra-rater agreement for electromyography, but a small agreement bias for range of motion measurements.
Background: The goal of this study was to assess the reliability of electromyography and range of motion measurements obtained using a knee exercise monitoring system. This device was developed to collect data on knee exercise activities.Methods: Twenty healthy individuals performed isotonic quadriceps exercises in this study. The vastus medialis surface electromyography (sEMG) and range of motion (ROM) of the knee were recorded during the exercise using the isotonic knee exercise monitoring device, the Mobi6-6b, and a video camera system. Each subject underwent a second measuring session at least 24 h after the first session. To determine reliability, the intraclass correlation coefficients (ICCs) and standard error of measurement (SEM) at the 95% confidence interval were calculated, and a Bland-Altman analysis was performed.Results: For inter-rater reliability, the ICCs of the mean absolute value (MAV) and root mean square (RMS) of sEMG were 0.73 (0.49, 0.86) and 0.79 (0.61, 0.89), respectively. ROM had an ICC of 0.93 (0.02, 0.98). The intra-rater reliability of the MAV of the sEMG was 0.89 (0.71, 0.96) and the intra-rater reliability of RMS of the sEMG was 0.88 (0.70, 0.95). The ROM between days had an intra-rater reliability of 0.82 (0.54, 0.93). The Bland-Altman analysis demonstrated no systematic bias in the MAV and RMS of sEMG, but revealed a small, systematic bias in ROM (-0.8311 degrees).Conclusion: For sEMG and range of motion measures, the isotonic knee exercise monitoring equipment revealed moderate to excellent inter- and intra-rater agreement. However, the confidence interval of ROM inter-rater reliability was quite large, indicating a small agreement bias; hence, the isotonic knee exercise monitor may not be suitable for measuring ROM. This isotonic knee exercise monitor could detect and collect information on a patient's exercise activity for the benefit of healthcare providers.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据