Performance Evaluation of the DxC 700 AU Chemistry in A1c Measurement

期刊

ANNALS OF LABORATORY MEDICINE
卷 43, 期 2, 页码 167-173

出版社

KOREAN SOC LABORATORY MEDICINE
DOI: 10.3343/alm.2023.43.2.167

关键词

Hemoglobin A1c; Fetal hemoglobin; Variant hemoglobin

向作者/读者索取更多资源

The performance of the DxC 700 AU immunoassay device in measuring HbA1c was evaluated. The device showed accuracy, precision, and linearity over a wide concentration range of HbA1c, but caution is needed in cases of high HbF and HbH concentrations.
Background: Accurate measurement of glycated hemoglobin (HbA1c) is crucial for a diabetes diagnosis and subsequent patient management. The detection method and presence of variant Hb can interfere with HbA1c measurements. We evaluated the HbA1c-measuring performance of the DxC 700 AU (Beckman Coulter, Brea, CA, USA) immunoassay based device in comparison with another immunoassay device and the reference method.Methods: A total of 120 normal and 14 variant Hb samples were analyzed using the Co bas c 513 (Roche Diagnostics, Mannheim, Germany) and DxC 700 AU analyzers. Variant Hb samples were also analyzed using the reference method, along with 20 normal samples. The accuracy, precision, linearity, and carryover were determined.Results: DxC 700 AU results strongly correlated with those of Cobas c 513 and exhibited accuracy in comparison with the reference method. The within-run, between-run, between day, and total imprecision (%CV) values for the low-and high-concentration control materials were below 2%. The results of DxC 700 AU were linear over a wide HbA1c range (3.39%-18.30%). Although DxC 700 AU performed well in the presence of variant Hb, the HbA1c concentration was underestimated in the presence of fetal Hb. The possibility of interference from a high HbH proportion could not be ruled out.Conclusions: The overall analytical performance of DxC 700 AU was acceptable. The device is accurate, precise, and linear over a wide HbA1c concentration range. Although DxC 700 AU results highly correlated with those of Cobas c 513, caution should be exercised in cases of high HbF and HbH concentrations.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据