Suitability of a Low-Cost Wearable Sensor to Assess Turning in Healthy Adults

期刊

SENSORS
卷 22, 期 23, 页码 -

出版社

MDPI
DOI: 10.3390/s22239322

关键词

inertial sensors; turning; validation; wearables

向作者/读者索取更多资源

This study aimed to evaluate the suitability of a low-cost wearable sensor device (AX6) for assessing turning outcomes. The results showed good to excellent agreement between AX6 and the reference device in different turning tasks. This suggests that AX6 can be a suitable and fit-for-purpose device for evaluating turning behavior.
Background: Turning is a complex measure of gait that accounts for over 50% of daily steps. Traditionally, turning has been measured in a research grade laboratory setting, however, there is demand for a low-cost and portable solution to measure turning using wearable technology. This study aimed to determine the suitability of a low-cost inertial sensor-based device (AX6, Axivity) to assess turning, by simultaneously capturing and comparing to a turn algorithm output from a previously validated reference inertial sensor-based device (Opal), in healthy young adults. Methodology: Thirty participants (aged 23.9 +/- 4.89 years) completed the following turning protocol wearing the AX6 and reference device: a turn course, a two-minute walk (including 180 degrees turns) and turning in place, alternating 360 degrees turn right and left. Both devices were attached at the lumbar spine, one Opal via a belt, and the AX6 via double sided tape attached directly to the skin. Turning measures included number of turns, average turn duration, angle, velocity, and jerk. Results: Agreement between the outcomes from the AX6 and reference device was good to excellent for all turn characteristics (all ICCs > 0.850) during the turning 360 degrees task. There was good agreement for all turn characteristics (all ICCs > 0.800) during the two-minute walk task, except for moderate agreement for turn angle (ICC 0.683). Agreement for turn outcomes was moderate to good during the turns course (ICCs range; 0.580 to 0.870). Conclusions: A low-cost wearable sensor, AX6, can be a suitable and fit-for-purpose device when used with validated algorithms for assessment of turning outcomes, particularly during continuous turning tasks. Future work needs to determine the suitability and validity of turning in aging and clinical cohorts within low-resource settings.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据