Parathyroid hormone reference intervals in adults using second- and third-generation assays

出版社

TAYLOR & FRANCIS LTD
DOI: 10.1080/00365513.2022.2164517

关键词

PTH; immunoassay; vitamin D; hypoparathyroidism; hyperparathyroidism

向作者/读者索取更多资源

Comparisons of different assay methods for parathyroid hormone (PTH) revealed variations in results and reference ranges. When switching from the second- to the third-generation PTH assay, lower levels were observed with the third-generation method, and method-specific reference ranges were established.
Parathyroid hormone (PTH) is a routine biochemical analysis, and it varies whether a second- or third-generation assay is used. Information on the levels obtained with different assays and evidence to substantiate local assay-specific reference ranges are important to inform clinical practice. Prior to a shift from the second- to the third-generation PTH assay (Cobas 8000, Roche Diagnostics) in the Department of Clinical Biochemistry, Aalborg University Hospital, Aalborg, Denmark, a total of 59 EDTA-plasma samples were collected for method comparison (Passing-Bablok). Furthermore, 120 EDTA-plasma samples were randomly obtained from adult blood donors and used for the establishment of reference intervals using the third-generation PTH assay (Cobas 8000, Roche Diagnostics) and two second-generation assays (Atellica, Siemens Healthineers; Alinity, Abbott Laboratories). Method comparison (Cobas 8000, Roche Diagnostics) showed lower levels with the third-generation (y) as compared to the second-generation assay (x) depending on the measurement range (PTH < 10 pmol/L: y = 0.8 (95% CI: 0.7; 0.9) x + 0.3 (95% CI: 0.2; 0.5), PTH >= 10 pmol/L: y = 0.6 (95% CI: 0.5; 0.6) x + 3.2 (95% CI: 1.1; 5.2)). Method-specific reference intervals (2.5 and 97.5 percentiles) after the exclusion of samples (n = 31) with 25-hydroxy-vitamin D below 50 nmol/L were: 1.8-8.5 pmol/L (second-generation, Atellica, Siemens Healthineers); 2.4-10.9 pmol/L (second-generation, Alinity, Abbott Laboratories), and 1.8-7.0 pmol/L (third-generation, Cobas 8000, Roche Diagnostics). PTH levels with second- and third-generation assays are not interchangeable. Clinicians should be informed when a laboratory assay is changed, and method-specific reference ranges are needed.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据