Hamstring muscle architecture assessed sonographically using wide field of view: A reliability study

期刊

PLOS ONE
卷 17, 期 11, 页码 -

出版社

PUBLIC LIBRARY SCIENCE
DOI: 10.1371/journal.pone.0277400

关键词

-

向作者/读者索取更多资源

The intra-rater reliability of static wide field of view ultrasound in quantifying the architectural characteristics of the hamstring muscles in male amateur athletes was assessed. The results showed that the architectural characteristics can be reliably quantified using static wide field of view ultrasound.
Purpose To assess the intra-rater reliability of static wide field of view ultrasound to quantify the architectural characteristics of the hamstring muscles. Methods Twenty amateur male athletes were recruited. Their left hamstring muscles were assessed using static wide field of view ultrasound on two separate occasions. Static ultrasound images were acquired by a single sonographer using a 92mm linear transducer. The architectural characteristics (muscle length, fascicle length, pennation angle and muscle thickness) from two distinct locations of the bicep femoris long head and semimembranosus were evaluated. Muscle length and thickness of the bicep femoris short head and semitendinosus muscle were also evaluated. Intraclass correlation coefficient analyses were performed to determine the intra-rater reliability of the performed measurements. Results Both muscle (intraclass correlation coefficient = 0.99; standard error of measurement = 4.3 to 6.6mm) and fascicle (intraclass correlation coefficient = 0.92 to 0.98; standard error measurement = 1.1 to 2.4mm) length were measured with excellent intra-rater reliability. Muscle thickness was measured with excellent reliability (intraclass correlation coefficient = 0.9 to 0.96; standard error of measurement = 0.91mm to 1.4mm) for all hamstring muscles except for the proximal segments of the bicep femoris short head (intraclass correlation coefficient = 0.85; standard error of measurement = 0.84mm) and semitendinosus (intraclass correlation coefficient = 0.88; standard error of measurement = 0.82mm), which were measured with good reliability. Pennation angle was measured with good reliability (intraclass correlation coefficient = 0.77 to 0.87; standard error of measurement = 1 to 1.6 degrees). Conclusion The architectural characteristics of the hamstring muscles of male amateur athletes can be reliably quantified using static wide field of view ultrasound.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据