The study presents a method to detect dead detector elements from flat field images using a convolutional neural network (CNN), aiming to provide a vendor-independent and potential correction algorithm-independent means for obtaining information that affects the image quality of diagnostic imaging systems. The results demonstrate the feasibility of the CNN method and suggest its potential application for a wide range of vendors and models in the evaluation of dead detectors.
BackgroundIndependent testing of image quality metrics is important to provide an unbiased determination of medical imaging performance. Due to the underreporting by vendors of dead detector elements, which are elements that do not function but may be corrected using information from surrounding pixels, the health of all imaging detector elements is infrequently reported and extraordinarily difficult to independently determine or verify through traditional means without vendor largesse. In many instances, dead detector data, or dead pixel maps, are only available at the discretion of the vendors, which renders these data inaccessible to many medical physicists. PurposeWe provide a mechanism to predict which detector elements are dead from a single flat field image and describe a convolutional neural network (CNN) to address this task. This technique provides a vendor-independent and potential correction algorithm-independent means for obtaining this information, which directly relates to the image quality of the diagnostic imaging system. MethodsA data set of 61 flat field images was obtained from two Varian on-board imaging (OBI) systems. These images were taken with a range of kVp and mAs settings and with differing levels of copper filtration in the beam path. The dead pixel maps for these detectors were also acquired and used to label the ground truth subimages-or small square images from the original, larger image-from the flat field images. This data set was then used to train, test, and validate five unique CNNs on the task of identifying dead detector elements from subimages derived from the flat field images. ResultsThe CNNs were validated with an average precision of 0.96, an average recall of 0.48, and an average F-1 score of 0.55 attributed to the scarcity and imbalanced nature of available data, and the difficulty in training neural networks using flat-field images. However, performance drastically improved when utilizing a subset of the data with a high dynamic range. ConclusionThis work shows the feasibility of using a CNN to detect dead detector elements from flat field images taken on a detector. Robust training of this type of algorithm could lead to a generalized model that may provide an independent evaluation of dead detectors for a wide range of vendors and models. Ultimately, this method can be a valuable tool for physicists performing quality assurance procedures on all digital diagnostic imaging systems, including X-ray, mammography, and fluoroscopy.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据