Detection of inflow obstruction in left ventricular assist devices by accelerometer: An in vitro study

期刊

MEDICAL ENGINEERING & PHYSICS
卷 110, 期 -, 页码 -

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.medengphy.2022.103917

关键词

Accelerometer; LVAD; Inflow obstructions; Pump vibrations; Detection

向作者/读者索取更多资源

The accelerometer detected LVAD inflow obstructions much better than PLVAD, potentially offering a monitoring modality for any pendulating obstructive inflow pathology in the future.
Inflow obstruction in left ventricular assist devices (LVAD) may lead to embolic stroke and pump malfunction. We investigated if an accelerometer detected graded LVAD inflow obstructions. Detection performances were compared to the current continuous surveillance routine based on the pump power consumption (PLVAD).In ten mock circuit experiments, four different-sized pendulating balloons obstructed HVADTM inflow conduits cross-section areas by 14%-75%. Nonharmonic amplitudes (NHA) of continuous signals from a triaxial accel-erometer attached to the LVAD were compared against single-point PLVAD values, using load and speed alter-ations as control interventions. We analyzed the NHA band power with a pairwise nonparametric statistical test. The detection performances were analyzed by receiver operating characteristics with areas under the curves (AUC).The NHA remained unaffected during load alterations. In contrast, NHA increased significantly from the 27% obstruction level (AUC >= 0.82), an effect amplified by increased pump speed. PLVAD did not change significantly below the maximal 75% obstruction level (AUC <= 0.36).In conclusion, NHA detected the inflow obstructions much better than PLVAD. The technique may provide a future monitoring modality of any pendulating obstructive inflow pathology.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据