4.6 Article

A high throughput solubility assay for drug discovery using microscale shake-flask and rapid UHPLC-UV-CLND quantification

期刊

出版社

ELSEVIER
DOI: 10.1016/j.jpba.2016.01.022

关键词

Rapid quantitative analysis; CLND-chemiluminescence nitrogen detector; HPLC hyphenated technique; Shake-flask solubility assay; Physicochemical properties

向作者/读者索取更多资源

The rapid determination of key physical properties of lead compounds is essential to the drug discovery process. Solubility is one of the most important properties since good solubility is needed not only for obtaining reliable in vitro and in vivo assay results in early discovery but also to ensure sufficient concentration of the drug being in circulation to get the desired therapeutic exposure at the target of interest. In order for medicinal chemists to tune solubility of lead compounds, a rapid assay is needed to provide solubility data that is accurate and predictive so that it can be reliably used for designing the next generation of compounds with improved properties. To ensure speed and data quality, we developed a high throughput solubility assay that utilizes a single calibration UHPLC-UV-CLND method and a 24 h shake-flask format for rapid quantification. A set of 46 model compounds was used to demonstrate that the method is accurate, reproducible and predictive. Here we present development of the assay, including evaluation of quantification method, filtration membranes, equilibrium times, DMSO concentrations, and buffer conditions. A comparison of thermodynamic solubility results to our high throughput 24 h shake-flask solubility assay results is also discussed. (C) 2016 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据