Stroke volume and cardiac output evaluation by electrical cardiometry: accuracy and reference nomograms in hemodynamically stable preterm neonates

期刊

JOURNAL OF PERINATOLOGY
卷 36, 期 9, 页码 748-752

出版社

NATURE PUBLISHING GROUP
DOI: 10.1038/jp.2016.65

关键词

-

向作者/读者索取更多资源

OBJECTIVE: To investigate the accuracy of electrical cardiometry (EC) to measure stroke volume (SV) and cardiac output (CO) and to provide gestational age (GA) and birth weight (BW)-based reference data for SV and CO in hemodynamically stable preterm neonates. STUDY DESIGN: Prospective observational blinded study. Paired measurements of SV and CO on stable preterm infants without any hemodynamic compromise were carried out using EC (SVEC) and echocardiography (SVECHO). RESULTS: Seventy-nine preterm neonates (mean GA: 31 +/- 3.2 weeks) were enrolled. A good correlation was found for SV (r = 0.743; P < 0.0001) and CO (r = 0.7; P < 0.0001) measured by EC and echocardiography. These correlations remained significant after adjusting for GA, patent ductus arteriosus and type of respiratory support (SV: St.beta = 0.48, P < 0.0001 and CO: st.beta = 0.69, P < 0.0001). Mean biases (and variabilities) were -1.1 (from 0.7 to -2.9) ml and -0.21 (from 0.15 to -0.55) I min(-1) for SV and CO, respectively. Local regression shows a tendency for EC to overestimate SV and CO especially at higher values (at about >2 ml and >0.41 min(-1), respectively). Coefficient of variation of SV was 48.9% and 52%, for EC and echocardiography. SV and CO rose with increasing GA and BW following an exponential equation (R-2>0.8). CONCLUSION: Measuring SV and CO with EC in hemodynamically stable preterm infants shows good correlation and variability similar to that of echocardiography. A trend to overestimation exists at highest values, but it is unlikely to be clinically significant. Reference GA and BW-based nomograms for SV and CO are provided.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据