4.2 Article

Rationale and design of the Lead Evaluation for Defibrillation and Reliability study: Safety and efficacy of a novel ICD lead design

期刊

JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY
卷 34, 期 2, 页码 257-267

出版社

WILEY
DOI: 10.1111/jce.15747

关键词

adaptive design; defibrillation lead; virtual patient modeling

向作者/读者索取更多资源

The LEADR Clinical Study aims to assess the long-term reliability of a novel defibrillator lead design using Bayesian methods and virtual patient data. The study will enroll 500 to 900 patients for 18 months to 3 years of follow-up, with multiple clinical assessments at different time points.
BackgroundImplantable cardioverter defibrillators (ICD) are indicated for primary and secondary prevention of sudden cardiac arrest. Despite enhancements in design and technologies, the ICD lead is the most vulnerable component of the ICD system and failure of ICD leads remains a significant clinical problem. A novel, small-diameter, lumenless, catheter-delivered, defibrillator lead was developed with the aim to improve long-term reliability. Methods and ResultsThe Lead Evaluation for Defibrillation and Reliability (LEADR) study is a multi-center, single-arm, Bayesian, adaptive design, pre-market interventional pivotal clinical study. Up to 60 study sites from around the world will participate in the study. Patients indicated for a de novo ICD will undergo defibrillation testing at implantation and clinical assessments at baseline, implant, pre-hospital discharge, 3 months, 6 months, and every 6 months thereafter until official study closure. Patients may be participating for a minimum of 18 months to approximately 3 years. Fracture-free survival will be evaluated using a Bayesian statistical method that incorporates both virtual patient data (combination of bench testing to failure with in-vivo use condition data) with clinical patients. The clinical subject sample size will be determined using decision rules for number of subject enrollments and follow-up time based upon the observed number of fractures at certain time points in the study. The adaptive study design will therefore result in a minimum of 500 and a maximum of 900 patients enrolled. ConclusionThe LEADR Clinical Study was designed to efficiently provide evidence for short- and long-term safety and efficacy of a novel lead design using Bayesian methods including a novel virtual patient approach.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据