4.6 Review

The Role of Dual-Mobility Components in Total Hip Arthroplasty

期刊

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.2106/JBJS.22.00185

关键词

-

向作者/读者索取更多资源

Dual mobility (DM) is a widely used option for total hip replacement, which features a small inner head within a larger outer polyethylene head. It aims to increase stability and range of motion in the hip joint. However, long-term data, especially for modular DM implants, is lacking, and potential issues include accelerated polyethylene wear, intraprosthetic dislocation, and modular backside fretting corrosion.
Dual mobility (DM) refers to a now widely available option for total hip articulation. DM implants feature a small inner head, a hard bearing, that connects via a taper fit onto the femoral trunnion. This head freely rotates but is encased inside a larger, outer polyethylene head that articulates with a smooth acetabular component. DM acetabular components are available in the form of a monoblock shell or as a liner that is impacted into a modular shell, providing a metal articulation for the polyethylene outer head. DM is designed to increase hip stability by providing the arthroplasty construct with a higher jump distance, head-to-neck ratio, and range of motion prior to impingement. The use of DM in total hip arthroplasty continues to increase in the United States for both primary and revision arthroplasty. Surgeons should be aware of the potential benefits and pitfalls. Long-term data are lacking, especially for modular DM implants. Points of concern include a potential for accelerated polyethylene wear, intraprosthetic dislocation, and modular backside fretting corrosion.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据