Dosimetric accuracy of the Convolution algorithm for Leksell Gamma Plan radiosurgery treatment planning: Evaluation in the presence of clinically relevant inhomogeneities

期刊

出版社

WILEY
DOI: 10.1002/acm2.13903

关键词

bone inhomogeneity; dose calculation accuracy; dose to medium; Gamma Plan Convolution; Leksell Gamma Knife

向作者/读者索取更多资源

In this study, the accuracy of the LGP-Convolution algorithm was evaluated in scenarios involving bone and/or air inhomogeneities. The results demonstrated excellent agreement between LGP-Convolution calculations and experimental as well as Monte Carlo results of dose to medium. It was confirmed that the LGP-Convolution algorithm reports the dose to medium, and the air inhomogeneity was found to have no clinical significance.
PurposeThe Leksell Gamma Plan Convolution algorithm (LGP-Convolution) has not been widely adopted. This mainly stems from the higher calculated beam-on times relative to the standard ray tracing-based LGP-TMR10 dose calculation algorithm. This study aims to evaluate the accuracy of the LGP-Convolution in scenarios where the treated lesions are in the vicinity of or encompassed by bone and/or air inhomogeneities. MethodsThe solid water dosimetry phantom provided by the vendor was modified to include bone and air inhomogeneities. Two treatment planning scenarios were investigated involving a single shot and multiple shots, respectively. Treatment planning and dose prescription were performed using the LGP-Convolution algorithm. Triple channel film dosimetry was performed using GafChromic EBT3 films calibrated in terms of absorbed dose to water in a Co-60 beam. Monte Carlo (MC) simulation dosimetry was also performed in the inhomogeneous experimental geometry using the EGSnrc MC platform and a previously validated sector-based phase-space source model. MC simulations were also employed to determine correction factors required for converting EBT3 measurements at points within the bone and air inhomogeneities from dose-to-water values to the corresponding dose to medium values. Results and ConclusionsEBT3 dose to medium correction factors ranged with field size (4, 8, or 16 mm) within 0.941-0.946 for bone and 0.745-0.749 for air inhomogeneities. An excellent agreement was found between the LGP-Convolution calculations with corresponding EBT3 and MC dose to medium results at all measurement points, except those located inside the air inhomogeneity. The latter is of no clinical importance and excluding them yielded gamma index passing rates of nearly 100% for 3% local dose difference and 1 mm distance-to-agreement criteria. The excellent agreement observed between LGP-Convolution calculations and film as well as MC results of dose to medium indicates that the latter is the quantity reported by the LGP-Convolution.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据