4.1 Article

Macrogeometric Differentiation of Dental Implant Primary Stability: An In Vitro Study

期刊

出版社

QUINTESSENCE PUBLISHING CO INC
DOI: 10.11607/jomi.9656

关键词

dental implant; in vitro; insertion torque; macrogeometry; primary stability

资金

  1. Dentsply Implants

向作者/读者索取更多资源

The purpose of this study was to evaluate an alternative cutting, progressive thread design to increase primary stability in low-density bone. The results showed that implants with cutting and progressive thread design achieved relatively high primary stability in low-density bone.
Purpose: To evaluate an alternative cutting, progressive thread design to increase primary stability in low-density bone. Materials and Methods: Four different implants with different macrogeometries (Nobel Biocare Active [NA] Internal RP implants [4.3 x 11.5 mm], Straumann BLX Roxolid RB implants [4.5 x 10 mm], Astra Tech Implant EV implants [4.2 x 11 mm], and PrimeTaper [PT 4.2 x 11 mm]) were placed in simulated osteotomies and extraction sockets in synthetic bone (Sawbones) according to the manufacturers' protocol. Insertion torque and ISQ values were measured using Implantmed Plus motor and Ostell IDX, respectively. Insertion time was recorded. Average values were calculated and compared using ANOVA and Tukey test. Results: Insertion torque (range: 5 to 44 Ncm) increased with increasing synthetic bone density for all implants. Different ISQ values in synthetic low-density bone were not observed in higher-density synthetic bone. Insertion torque of all implants was reduced when implants were placed in simulated sockets compared to simulated osteotomies. In both low- density and higher-density synthetic bone, the primary stability of PrimeTaper implants with cutting and progressive thread design was equivalent to that of the Nobel Biocare NobelActive implant with compressive thread design and greater than the BLX implant with compressive thread design. Conclusion: Different implant macrogeometries obtain relatively high primary stability in low-density bone when measured by ISQ. Double-thread implant designs reduce insertion times in higher-density bone. A cutting and progressive compressing thread design provides density-sensing performance compared to aggressive condensing thread designs. This macrogeometry can achieve high primary stability associated with modest insertion torque compared to aggressive threaded implant designs known to attain the highest insertion torque. The presence of multiple cutting threads may offer advantages in obtaining primary stability in low-density bone.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据