4.4 Article

Comparison of the clinical performance of a hybrid Alba 4D and the AMC-4 locoregional hyperthermia systems

期刊

INTERNATIONAL JOURNAL OF HYPERTHERMIA
卷 39, 期 1, 页码 1408-1414

出版社

TAYLOR & FRANCIS LTD
DOI: 10.1080/02656736.2022.2140841

关键词

Hyperthermia; locoregional hyperthermia systems; clinical performance; AMC-4 system; Alba 4D system

向作者/读者索取更多资源

The study demonstrates similar clinical performance between the hybrid Alba 4D and the AMC-4 locoregional hyperthermia systems, indicating comparable treatment outcomes for patients.
Objective: The in-house developed 70MHz AMC-4 locoregional hyperthermia system has been in clinical use since 1984. This device was recently commercialized as the Alba 4D (Medlogix (R), Rome, Italy), with a similar geometrical 4-waveguide design. At the time of this study a hybrid Alba 4D was installed at our center, which incorporated elements of the AMC-4. This study aims to compare clinical performance of both devices. Methods: During one year after clinical acceptance of the hybrid Alba 4D, both devices were used for treatment delivery in patients scheduled for locoregional hyperthermia. Each patient started with the AMC4, next sessions were allocated to either device. Possible differences between Alba 4D and AMC-4 sessions in power, achieved temperature T0, T10, T50, T90, T100, treatment time and complaints per session, were evaluated using linear mixed models (LMMs) for repeated measures with patient as random effect. Results: From March 2018 to April 2019, eleven patients with cervical, pancreatic, vaginal carcinoma and uterine leiomyosarcoma received 27 locoregional hyperthermia sessions with the Alba 4D and 34 sessions with the AMC-4. Median number of sessions per patient was 5 (range 3-13). Treatment results for both devices were not significantly different: T50 was 40.5 +/- 1.0 degrees C vs. 40.8 +/- 0.7 degrees C, applied power was 500 +/- 79 W vs. 526 +/- 108W, for the Alba 4D vs. AMC-4, respectively. Conclusion: Results of the first patients treated with the hybrid Alba 4D demonstrated comparable clinical performance of the Alba 4D and AMC-4 locoregional hyperthermia systems, and both devices are expected to yield similar favorable clinical results.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据