Combination of current and new electrocardiographic-based criteria: a novel score for the discrimination of left bundle branch capture

期刊

EUROPACE
卷 25, 期 3, 页码 1051-1059

出版社

OXFORD UNIV PRESS
DOI: 10.1093/europace/euac276

关键词

Electrocardiographic-based criteria; Left bundle branch capture; Left ventricular septal capture; Conduction system pacing

向作者/读者索取更多资源

This study found that the combination of multiple ECG criteria, including V6-RWPT, aVL-RWPT, and V6-V1 interpeak interval, can better differentiate LBB capture from LVS capture. These criteria demonstrated better diagnostic performance than isolated measurements.
Aims Most of the criteria used to diagnose direct capture of the left bundle branch (LBB) have never been validated in an external sample. We hypothesized that lead aVL might add relevant information, and the combination of several electrocardiograph (ECG)-based criteria might discriminate better LBB capture from left ventricular septal (LVS) capture, than each criterion separately. Methods and results Single-centre study involving all consecutive patients who received LBB area pacing. LBB capture was defined according to QRS morphology transition criteria during decremental pacing. Multivariate logistic regression analysis was performed to develop a predictive score for LBB capture. A total of 71 patients with confirmed LBB capture were analysed. The optimal cut-off values of R wave peak time (RWPT) in lead V6 (V6-RWPT) and V6-V1 interpeak interval for the discrimination of LBB capture were <83 ms and >= 33 ms, respectively. The RWPT in lead aVL (aVL-RWPT) showed a good discrimination power for the differential diagnosis of LBB capture and LVS capture. The optimal value for aVL-RWPT was 79 ms [sensitivity (SN) and specificity (SP) of 71.2% and 88.4%, respectively]. A new score, with a good diagnostic performance (area under the curve of 0.976), was constructed gathering the information from V6-RWPT, aVL-RWPT, and V6-V1 interpeak interval. The optimal score of 3 points showed a SN and SP of 89.2% and 100%, respectively for the differentiation of LBB capture. Conclusions ECG-based criteria are useful to confirm the capture of the LBB. The combination of V6-RWPT, aVL-RWPT, and V6-V1 interpeak interval values demonstrated better diagnostic performance than isolated measurements.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据