4.3 Article

Clinical Evaluation of Computer-Aided Colorectal Neoplasia Detection Using a Novel Endoscopic Artificial Intelligence: A Single-Center Randomized Controlled Trial

期刊

DIGESTION
卷 104, 期 3, 页码 193-201

出版社

KARGER
DOI: 10.1159/000528085

关键词

Artificial intelligence; Computer-aided detection; Colon polyps; Adenoma detection rate; Adenoma miss rate

向作者/读者索取更多资源

This study evaluated the clinical performance of a new endoscopic artificial intelligence system for computer-aided detection (CADe) of colonic adenomas. The CADe group showed a 11.8% higher adenoma detection rate (ADR) compared to the control group. Additionally, there was no need to extend the examination time or request the assistance of additional medical staff to achieve this improved effectiveness.
Introduction: Computer-aided diagnostic systems are emerging in the field of gastrointestinal endoscopy. In this study, we assessed the clinical performance of the computer-aided detection (CADe) of colonic adenomas using a new endoscopic artificial intelligence system. Methods: This was a single-center prospective randomized study including 415 participants allocated into the CADe group (n = 207) and control group (n = 208). All endoscopic examinations were performed by experienced endoscopists. The performance of the CADe was assessed based on the adenoma detection rate (ADR). Additionally, we compared the adenoma miss rate for the rectosigmoid colon (AMRrs) between the groups. Results: The basic demographic and procedural characteristics of the CADe and control groups were as follows: mean age, 54.9 and 55.9 years; male sex, 73.9% and 69.7% of participants; and mean withdrawal time, 411.8 and 399.0 s, respectively. The ADR was 59.4% in the CADe group and 47.6% in the control group (p = 0.018). The AMRrs was 11.9% in the CADe group and 26.0% in the control group (p = 0.037). Conclusion: The colonoscopy with the CADe system yielded an 11.8% higher ADR than that performed by experienced endoscopists alone. Moreover, there was no need to extend the examination time or request the assistance of additional medical staff to achieve this improved effectiveness. We believe that the novel CADe system can lead to considerable advances in colorectal cancer diagnosis.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据