Small globes and pocket heat stress meters for WBGT and PHS evaluations. A critical analysis under controlled conditions

期刊

BUILDING AND ENVIRONMENT
卷 226, 期 -, 页码 -

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.buildenv.2022.109781

关键词

Globe temperature; Mean radiant temperature; Heat stress meters; WBGT; PHS

向作者/读者索取更多资源

This paper quantifies the effects of using small globes to assess extreme hot thermal environments. Results show that small globes underestimate WBGT values by 1°C, but this can be corrected using specific correction coefficients. However, pocket WBGT meters based on small globes and psychrometric sensors underestimate WBGT by more than 10°C, especially at low air velocity and high radiative loads. Additionally, the underestimation of mean radiant temperature measured through small globes leads to an underestimation of allowable exposure times predicted by the PHS model.
This paper is aimed at quantifying the effects of the well-known criticalities related to the use of small globes in the assessment of extreme hot thermal environments. Based on the experimental measurements carried out through a special apparatus able to reproduce a black enclosure under controlled conditions, the analysis has been focused on both screening (WBGT) and rational (PHS) assessment methods. Obtained results showed the underestimation of the WBGT values within 1 degrees C using small globes for measuring the WBGT index according to the Standard ISO 7243 and no relevant errors occur also in the presence of high radiative loads. These errors can be even eliminated if specific correction coefficients for the heat transfer by convection are used. On the other hand, pocket WBGT meters based upon small globes and psychrometric (shielded) sensors - instead of the wet wick - are affected by an unacceptable underestimation (by more than 10 degrees C) of the WBGT that is particularly enhanced at low air velocity and high radiative loads. Finally, the systematic underestimation of the mean radiant temperature measured through small globes can lead the allowable exposure times predicted by the PHS model to be underestimated by more than 5 h when a 38 mm table-tennis ball is used.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据