Liquid crystal elastomer based dynamic device for urethral support: Potential treatment for stress urinary incontinence

期刊

BIOMATERIALS
卷 292, 期 -, 页码 -

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.biomaterials.2022.121912

关键词

Liquid crystal elastomers; Artificial muscle; Direct ink writing; Stress urinary incontinence

向作者/读者索取更多资源

Stress urinary incontinence (SUI) is the involuntary loss of urine due to increased intra-abdominal pressure. This study reports the fabrication of a novel device based on liquid crystal elastomers (LCEs) that can change shape in response to temperature increase induced by transcutaneous IR light. The device was tested in a scar tissue phantom model and in an in vitro urinary tract model, and was acutely implanted and tested in female rabbits.
Stress urinary incontinence (SUI) is characterized by the involuntary loss of urine due to increased intra-abdominal pressure during coughing, sneezing, or exercising. SUI affects 20-40% of the female population and is exacerbated by aging. Severe SUI is commonly treated with surgical implantation of an autologous or a synthetic sling underneath the urethra for support. These slings, however, are static, and their tension cannot be non-invasively adjusted, if needed, after implantation. This study reports the fabrication of a novel device based on liquid crystal elastomers (LCEs) capable of changing shape in response to temperature increase induced by transcutaneous IR light. The shape change of the LCE-based device was characterized in a scar tissue phantom model. An in vitro urinary tract model was designed to study the efficacy of the LCE-based device to support continence and adjust sling tension with IR illumination. Finally, the device was acutely implanted and tested for induced tension changes in female multiparous New Zealand white rabbits. The LCE device achieved 5.6% +/- 1.1% actuation when embedded in an agar gel with an elastic modulus of 100 kPa. The corresponding device temperature was 44.9 degrees C +/- 0.4 degrees C, and the surrounding agar temperature stayed at 42.1 degrees C +/- 0.4 degrees C. Leaking time in the in vitro urinary tract model significantly decreased (p < 0.0001) when an LCE-based cuff was sutured around the model urethra from 5.2min +/- 1min to 2min +/- 0.5min when the cuff was illuminated with IR light. Normalized leak point force (LPF) increased significantly (p = 0.01) with the implantation of an LCE-CB cuff around the bladder neck of multiparous rabbits. It decreased significantly (p = 0.023) when the device was actuated via IR light illumination. These results demonstrate that LCE material could be used to fabricate a dynamic device for treating SUI in women.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据