This study analyzed Korean and foreign systems and proposed a method to determine the replacement time for high-risk medical devices, revealing a lack of specific regulations regarding the life cycle of such devices. The study also identified important factors for device replacement evaluation, including year of introduction, repair cost, component discontinuation, and multiple failures.
In this study, we analyzed Korean and foreign systems, focusing on high-risk medical devices that urgently need to be managed, and we present an life cycle calculation method for determining replacement time. A literature review was conducted to confirm the regulations of the medical device management system and life cycle by country, and a case analysis was performed to verify the replacement evaluation criteria of actual medical institutions. In addition, durability data from the Public Procurement Service, American Hospital Association, and Samsung Medical Center were used to calculate the life cycle of high-risk medical devices. The analysis showed that in the case of Korean and foreign medical device regulatory agencies, there were no specific life cycle regulations for high-risk medical devices. In addition, the important items in the medical device replacement evaluation were found to be the year of introduction, repair cost, component discontinuation, and several failures. On calculating the life cycle of high-risk medical devices revealed that the replacement time is 13 years for anesthesia machines, 14 years for defibrillators, 16 years for heart-lung machines, and 13 years for ventilators. To introduce a uniform medical device replacement standard and life cycle calculation method, the government will need to reorganize the medical device replacement laws and systems. In addition, in the case of medical institutions, it is necessary to secure patient safety by using expert groups to prepare specific life cycle standards that consider the characteristics of medical devices.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据