Offset Compensation in Resistive Stretch Sensors Using Low-Frequency Feedback Topology

期刊

ELECTRONICS
卷 11, 期 19, 页码 -

出版社

MDPI
DOI: 10.3390/electronics11193158

关键词

sleep apnea; stretch sensor; signal conditioning; offset attenuation

资金

  1. AGH University of Science and Technology in Krakow [16.16.230.434]

向作者/读者索取更多资源

Respiration monitoring systems play a crucial role in healthcare and fitness. In this paper, a novel signal conditioning circuit for a resistive respiration sensor is proposed, which addresses some of the issues associated with existing sensors. The results indicate that resistive stretch sensors, combined with low-frequency feedback topology, offer promising prospects for future respiration monitoring devices.
Respiration monitoring systems play an important role in healthcare and fitness. For this purpose, resistive stretch sensors are frequently used, which are cheap and simple in operation. However, they are not free from drawbacks. Varying offset due to patient movement, low signal amplitude, as well as susceptibility to interference, can all pose serious challenges. In this paper, a novel signal conditioning circuit for a resistive respiration sensor is proposed that alleviates some of the above problems. Namely, the proposed low-frequency feedback topology improves the dynamic range by offset compensation, sustaining a high signal amplification. Further advantages of the new configuration are the phase shift of 0.5 degrees in the band of interest and higher gain for the respiration signal than for the offset. The topology was proved to correctly represent signal amplitude changes, as well as to be able to sample human respiration in the home environment. However, the circuit shows some nonlinear behavior around resistance discontinuity points-settling time after body position change of the patient, which can be as long as 40 s. The circuit was tested both in bench tests and in the prototype of a respiratory polygraphy device during actual sleep apnea examinations. The results indicate that resistive stretch sensors, along with low-frequency feedback topology, are a promising development path for future respiration monitoring devices.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据