Impedance evaluation of textile electrodes for EEG measurements

期刊

TEXTILE RESEARCH JOURNAL
卷 93, 期 7-8, 页码 1878-1888

出版社

SAGE PUBLICATIONS LTD
DOI: 10.1177/00405175221135131

关键词

Conductive textiles; EEG; textile electrodes; wearable technology

向作者/读者索取更多资源

This paper evaluated the impedance of textile electrodes for electroencephalography measurements. The study found that copper-based textile electrodes were more suitable than silver-based textile electrodes and comparable to commercial electrodes in terms of impedance.
In this paper, we evaluated the impedance of textile electrodes for electroencephalography measurements. We compared the commercially available standard dry silver/silver chloride dry electrodes, copper-based, and silver-based textile electrodes by measuring impedance on the skin. First, on dry skin, the impedance range of the electrodes was 93-350 k omega at 0 min; after 60 min, the range was 20-56 k omega. The commercial electrode showed the lowest impedance of 20 k omega, and the silver-based textile electrode showed the highest impedance of 56 k omega in dry skin conditions after 60 min. When tap water was applied as an electrolyte, the impedance range was 13-15 k omega at 0 min; after 60 min, the range was 17-22 k omega for all three types of electrodes. In this case, the copper-based textile electrode showed the lowest impedance of 17 k omega and the commercial electrode showed the highest impedance of 22 k omega after 60 min. When saline water was applied as an electrolyte, at 0 min, the impedance range was 8-11 k omega; after 60 min, the range became 11-12 k omega for all three types of electrodes. Thus, all electrodes showed similar results. It was concluded that the copper-based textile electrodes were more suitable than the silver-based textile electrodes and comparable to commercial electrodes regarding impedance in all three skin conditions.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据