4.4 Article

Design and transmission modeling of a soft elbow exosuit using double artificial tendon system

出版社

SAGE PUBLICATIONS LTD
DOI: 10.1177/09544062221129824

关键词

Soft exosuit; elbow joint; double artificial tendon system; transmission modeling; performance evaluation

向作者/读者索取更多资源

In order to assist in the rehabilitation of movement disorders, a lightweight, soft exosuit for elbow flexion and extension is developed based on muscle length relationship and the Hill-type muscle model. The prototype of the exosuit is capable and effective in helping the elbow joint.
Daily life would be affected if the upper extremity is injured, especially for people with spinal cord injury, stroke and other neuromuscular disorders. In order to assist in the rehabilitation of these movement disorders, exoskeleton robots are developed. Based on the muscular length relationship and the Hill-type muscle model, a lightweight, soft exosuit is designed for elbow flexion and extension. The exosuit is composed of an actuation box and a wearing sleeve. The power is transmitted from the actuation box to the sleeve utilizing an artificial tendon transmission system. The detailed mechanism design is introduced, and the mathematical double artificial tendon transmission model is developed. Experiments and related simulations are followed to validate the model. The prototype of the soft exosuit is tested on a male subject, and the results indicate that the soft exosuit is capable and effective in helping the elbow joint with flexion and extension.