Absorbed-dose-to-water measurement using alanine in ultra-high-pulse-dose-rate electron beams

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 67, 期 20, 页码 -

出版社

IOP Publishing Ltd
DOI: 10.1088/1361-6560/ac950b

关键词

alanine dosimetry; UHPDR; Monte Carlo; electron beam; integrating current transformer (ICT)

资金

  1. EMPIR programme [18HLT04 UHDpulse]
  2. European Union

向作者/读者索取更多资源

The aim of this study was to evaluate the dose response of PTB's alanine dosimeter system in UHPDR electron beams, with results showing a good precision of the system in the dose range of 0.15 Gy to 6.2 Gy per pulse. The Monte Carlo simulations were used to determine the beam quality conversion and correction factors needed for alanine dosimetry.
Objective. The aim of the presented study is to evaluate the dose response of the PTB's secondary standard system, which is based on alanine and electron spin resonance (ESR) spectroscopy measurement, in ultra-high-pulse-dose-rate (UHPDR) electron beams. Approach. The alanine dosimeter system was evaluated in the PTB's UHPDR electron beams (20 MeV) in a range of 0.15-6.2 Gy per pulse. The relationship between the obtained absorbed dose to water per pulse and the in-beamline charge measurement of the electron pulses acquired using an integrating current transformer (ICT) was evaluated. Monte Carlo simulations were used to determine the beam quality conversion and correction factors required to perform alanine dosimetry. Main results. The beam quality conversion factor from the reference quality Co-60 to 20 MeV obtained by Monte Carlo simulation, 1.010(1), was found to be within the standard uncertainty of the consensus value, 1.014(5). The dose-to-water relative standard uncertainty was determined to be 0.68% in PTB's UHPDR electron beams. Significance. In this investigation, the dose-response of the PTB's alanine dosimeter system was evaluated in a range of dose per pulse between 0.15 Gy and 6.2 Gy and no evidence of dose-response dependency of the PTB's secondary standard system based on alanine was observed. The alanine/ESR system was shown to be a precise dosimetry system for evaluating absorbed dose to water in UHPDR electron beams.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据