4.4 Article

Using barium as an internal radioprotective shield for pregnant patients undergoing CT pulmonary angiography: A retrospective study

期刊

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.ejmp.2022.08.014

关键词

Pregnancy; CT; Pulmonary embolism; Radiation protection

向作者/读者索取更多资源

The purpose of this retrospective study was to assess the effect of barium sulfate contrast medium on the radiation dose and diagnostic quality of CT pulmonary angiography (CTPA) in pregnant patients. The results indicated that the use of barium sulfate as an internal radioprotective shield produced counterproductive artifacts and tube current increments, limiting the evaluation of the lung and subsegmental lower lobe pulmonary arteries.
Purpose: The purpose of our retrospective study was to assess the effect of barium sulfate contrast medium on radiation dose and diagnostic quality of CT Pulmonary Angiography (CTPA) in an in-vivo study of pregnant patients. Methods: Our retrospective study included 33 pregnant patients who underwent CTPA to exclude pulmonary embolism. The patients received oral 40% w/v barium solution just prior to the acquisition of their planning radiograph. All CTPA were performed on 64-slice, single-source CT scanners with AEC with noise index = 28.62-31.64 and the allowed mA range of 100-450. However, only 5/33 patients had mA modulation (AEC 100-450 mA range), while 28/33 patients had mA maxed out at the set maximum mA of 450 over the entire scan range. We recorded CTDIvol (mGy), DLP (mGy.cm) and scan length. The same information was recorded in weight-and scanner-matched, non-pregnant patients. Statistical tests included descriptive data (median and interquartile range) and Mann-Whitney test. Results: There were no significant differences in CTDIvol and DLP between the barium and control group patients (p > 0.1). The median mA below the diaphragm was significantly higher in each patient with barium compared to the weight and scanner-matched patient without barium. Evaluation of lung and subsegmental lower lobe pulmonary arteries was limited in 85% barium group. Due to thin prospective section thickness (1.25 mm), most patients were scanned at maximum allowed mA for AEC. Conclusion: Use of AEC with thick barium in pregnant patients undergoing CTPA as an internal radioprotective shield produces counterproductive artifacts and tube current increments.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据