4.3 Article

The performance of a home-made 300 kV AMS system at CIAE

出版社

ELSEVIER
DOI: 10.1016/j.nimb.2022.07.008

关键词

Accelerator mass spectrometry; 300 kV; Home-made; I-129; Pu-239

资金

  1. National Natural Science Foundation of China [11875326]
  2. Continuous Basic Scientific Research Project [WDJC-2019-04]

向作者/读者索取更多资源

The China Institute of Atomic Energy has successfully developed a homemade 300 kV AMS system for heavy isotopes measurement, achieving good transmission efficiency and uncertainty in the measurement of I-129 and Pu-239.
In order to develop a AMS system for heavy isotopes measurement at China Institute of Atomic Energy (CIAE), a home-made 300 kV AMS system has been established. The measurement methods for I-129 and Pu-239 have been developed with the system. The transmission efficiency of 65 % for I-129 and 25 % for Uranium have been obtained with this system. The I-129 measurements with uncertainty of 2.5 % and sensitivity of I-129/I-127 = 2 x 10(-14) are achieved. The uncertainty of 5 % and detection limitation of 0.5 fg are obtained for Pu-239 measurements.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据