Influence of Pseudopotential Well Depth on the Performance of Hybrid Linear Ion Trap/Time-of-Flight Mass Spectrometry Driven by Square and Sinusoidal Waveform

期刊

出版社

AMER CHEMICAL SOC
DOI: 10.1021/jasms.2c00142

关键词

-

资金

  1. National Natural Science Foundation of China [21927804, 22027804, 22174142]
  2. Scientific Instrument Developing Project of the Chinese Academy of Sciences [ZDKYYQ20210005]
  3. Dalian High-Level Talent Innovation and Entrepreneurship Project [2020RQ097, 2020RQ095]
  4. Dalian Institute of Chemical Physics [DICP I202102]

向作者/读者索取更多资源

This paper investigates the differences in mass range, resolution, sensitivity, and ion fragmentation of LIT/TOFMS driven by square and sinusoidal waveforms. It is found that the square mode has a wider mass range, better mass resolution, and higher sensitivity compared to the sinusoidal mode. Additionally, the square mode exhibits softer behavior and less ion fragmentation. Therefore, LIT/TOFMS driven by a square waveform shows greater potential in complex matrix systems with a wide mass range.
Quadrupole devices are widely used as ion analyzers and ion guides, which are usually driven by a sinusoidal waveform. Recently, these devices driven by a digital waveform have attracted much attention. Driven by different rf waveforms, the pseudopotential well depth of quadrupole devices is significantly different, which will affect their performance. However, there is a lack of comprehensive view on the performance differences of quadrupole devices or hybrid instrument driven by different rf waveforms. In this paper, the differences in mass range, resolution, sensitivity, and ion fragmentation of LIT/TOFMS driven by square and sinusoidal waveforms were investigated. Compared with the sinusoidal mode, the square mode had a wider mass range and better mass resolution. Toluene was used to study the difference in sensitivity and ion fragmentation between the two modes. The results showed that the sensitivity of the square mode was 1.7-2.5 times that of the sinusoidal mode at different ion densities. According to the sensitivity differences, it was inferred that the ion-trapping efficiency and ion capacity of LIT in the square mode were 2.0 times and 1.7 times those in sinusoidal mode. It was found that the square mode behaved softer and presented less fragmentation, and the proportion of fragment ions in the sinusoidal mode can reach 3.2 times that in the square mode at high ion density. Therefore, LIT/TOFMS driven by a square waveform shows greater potential in the application of the complex matrix system with a wide mass range.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据