4.7 Article

Creatinine-Detecting Laser Diode-Induced Fluorescence Detection System

期刊

IEEE SENSORS JOURNAL
卷 22, 期 18, 页码 17784-17790

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/JSEN.2022.3196629

关键词

3.5-dinitrobenzoic acid (DNB); creatinine; D-lactate; fluorescence

资金

  1. Ministry of Science and Technology (MOST) of Taiwan [110-2637-E-027-002, 109-2637-E-027-002]

向作者/读者索取更多资源

A novel fluorescence-based bioassay for rapid and cost-effective detection of rat creatinine concentration was developed. The fluorescence intensity was used to calculate the creatinine concentration, and a fluorescence sensing system was established.
Under highly alkaline conditions, creatinine reacts with 3.5-dinitrobenzoic acid (DNB) to produce a product that can absorb ultraviolet light at a wavelength of 400 nm and emit fluorescence at a wavelength of 491 nm. Thus, creatinine concentration can be calculated using fluorescence intensity. A novel fluorescence-based bioassay for rat creatinine that is capable of rapid detection at a low cost was developed in this study. A fluorescence sensing system was established using a micro-spectrometer, a 3-D translation stage, and a laser diode (LD) to measure various concentrations of serum creatinine. The measurable range was 12.5- 300 mu mol/L ( R-2 = 0.996), and the limit of detection (LOD) was 5 mu mol/L. The intraday precision and accuracy ranges were 7.33%-10.11% and 81.82%-99.75%, respectively, while the interday precision and accuracy ranges were 3.08%-9.39% and 86.75%-87.85%, respectively. Additionally, the test results for rat serum were compared with those from the National Laboratory Animal Center of National Applied Research Laboratories in Taiwan, and the correlation coefficient was determined to be 0.80.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据