4.3 Article

Methods for anti-factor VIII antibody levels in haemophilia A patients - validation of a multiplex immunoassay and comparability with assays measuring non-neutralising and neutralising antibodies (inhibitors)

期刊

HAEMOPHILIA
卷 -, 期 -, 页码 -

出版社

WILEY
DOI: 10.1111/hae.14669

关键词

antibodies; factor VIII; haemophilia A; inhibitors; methods

向作者/读者索取更多资源

The xFLI method, based on Luminex xMAP fluorescence-immunoassay, is more sensitive and reproducible than ELISA and CBA for detecting anti-FVIII antibodies in patients with haemophilia A. It can detect both functional inhibitors and non-neutralising antibodies accurately. The overall concordance between xFLI and ELISA was 82.1% and xFLI and CBA was 77.9%.
Introduction The development of neutralising (inhibitors) and non-neutralising antibodies (NNAs) is a complication to factor replacement therapy in haemophilia. The diagnostic methods available lack standardisation, have high inter-laboratory variation, and false-negative as well as false-positive results may affect treatment. Both functional inhibitors and NNAs may be detected with higher reproducibility, sensitivity and specificity using the immunological Luminex xMAP-based fluorescence-immunoassay (xFLI). Aim Validation of our xFLI and comparability with enzyme-linked immunosorbent assay (ELISA) and chromogenic Nijmegen-Bethesda assay (CBA) for anti-FVIII antibodies in haemophilia A (HA) patients. Methods The xFLI method was developed with full-length and B-domain deleted factor coupled to magnetic beads, optimised and validated for performance characteristics. Comparability with ELISA and CBA was evaluated in HA patient samples (n = 112), serial samples in six inhibitor patients and reference interval and decision-limits in healthy donors (n = 44). Results The intra- and inter-assay precision (CV%) for the xFLI method was below 6% and detection limit (LLOQ) .084 ng/mL (NovoEight). All ELISA-positive samples were positive with either Advate or NovoEight. Additionally, 10.7%-14.3% were xFLI-positive and ELISA-negative. All but one CBA-positive sample was above 3SD with xFLI; one was between 2 and 3SD. 29.1% were xFLI-positive and CBA negative. The overall concordance between xFLI and ELISA was 82.1% and xFLI and CBA 77.9%. Conclusion The anti-FVIII antibody xFLI method is adaptable to clinical practice and more sensitive and reproducible than ELISA and CBA. Actual NNA titers are determined to both full-length and B-domain deleted FVIII. The xFLI is thus valuable for confirmation of all anti-FVIII antibodies.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据