4.7 Article

Simultaneous extraction/derivatization for the analysis of sulfite by capillary electrophoresis: A high-throughput reference method to meet the demand of seafood inspection

期刊

FOOD RESEARCH INTERNATIONAL
卷 161, 期 -, 页码 -

出版社

ELSEVIER
DOI: 10.1016/j.foodres.2022.111780

关键词

Sulfite; Capillary zone electrophoresis; Food analysis; Derivatization; Preservatives; Shrimp

资金

  1. National Council for Scientific and Technological Development (CNPq) [142228/2017-6]
  2. INCT Catalysis

向作者/读者索取更多资源

This study developed an innovative method for sulfite inspection in seafood using capillary zone electrophoresis. The method is simple, fast, and allows for simultaneous extraction and deriva-tization of sulfite, suitable for high throughput analysis. The method showed selectivity and specificity when compared to liquid chromatography coupled to mass spectrometry. The results met the required parameters for metrological accreditation.
Sulfite is widely used to prevent enzymatic browning in shrimp and lobster processing. However, sulfite may cause allergy in sensitive consumers. Thus, regulatory agencies set limits for its use. Sulfite is usually controlled by the normalized Monier-Williams (MW) titrimetric method that allows a limited number of samples to be analyzed. This manuscript consolidates an innovative method for sulfite inspection in seafood by capillary zone electrophoresis with diode array detector (CZE-DAD). A simple, fast, and simultaneous extraction and deriva-tization method was developed to provide high throughput for analytical routine. The high instability of the sulfite was suppressed by its derivatization with formaldehyde producing hydroxymethylsulfonate. The evalu-ation of its analytical performance yielded excellent results in compliance with the strict parameters required for metrological accreditation. The CZE-DAD method was selective and specific when submitted to confirmatory evaluations by liquid chromatography coupled to mass spectrometry. The limit of detection (3.50 mg kg(-1)), limit of quantitation (11.7 mg kg( -1)) and recoveries (99-103%) were adequate for sample analysis. The measurement uncertainty was estimated by the propagation of errors and experimental standard uncertainties (precision, accuracy, and analytical curves) and type B uncertainties from traceable measurement instruments. The low relative uncertainty (10%) and the adequate reproducibility demonstrated method suitability. The CZE-DAD results were compared to the MW method through the respective expanded standard uncertainties and normalized error. This new method is promising to be used in seafood inspection and continuous laboratory evaluations using instrumentation not very expensive to acquire and maintain.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据