Repeatability, reproducibility and agreement between three different diagnostic imaging platforms for tear film evaluation of normal and dry eye disease

期刊

EYE
卷 37, 期 10, 页码 2042-2047

出版社

SPRINGERNATURE
DOI: 10.1038/s41433-022-02281-2

关键词

-

向作者/读者索取更多资源

This study assessed the repeatability and reliability of different commercially available diagnostic platforms in the objective assessment of tear film parameters and found that they cannot be used interchangeably. There were significant differences in the measurement values of tear film parameters between different instruments, and patients with dry eye disease had higher variability in tear film parameters.
Purpose To assess the repeatability and reliability of different commercially available diagnostic platforms in the objective assessment of tear film parameters and if there exists any agreement between them. Methods Thirty healthy individuals (N = 60 eyes) and fifteen DED patients (N = 30 eyes) had their tear film parameters (Lipid layer thickness (LLT), Tear meniscus height (TMH), Non-invasive tear break up time (NIBUT)) assessed using three instruments - LipiView (R) II, IDRA ocular surface analyser (IDRA-OSA) and Oculus keratograph 5 M (K5M). Bland-Altman analysis and linear mixed effects modelling & Generalized Linear Hypothesis Test were used for analysis and coefficient of variation (CoV). Results There is poor repeatability but good reproducibility of LLT values measured with Lipiview, or IDRA. NIBUT using K5M & IDRA-OSA shows good repeatability and reproducibility in control group but poor repeatability in DED patients. TMH values obtained with K5M or IDRA-OSA had poor repeatability with high CoV. Between two observers, good reproducibility is observed for TMH and NIBUT values using both K5M & IDRA-OSA but not for LLT values. Between instruments, all the measurements (LLT, NIBUT and TMH) were significantly different on Bland Altman analysis. Conclusion No two dry eye diagnostic platforms can be used interchangeably and non-invasive tear film values should be interpreted keeping in mind the individual machine's variability. The high coefficient of variation in DED patients compared to normal reflects inherent variability in tear film irrespective of the device used.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据