A 3D fluid structure interaction computational model was developed to simulate the interaction between bileaflet mitral valve and blood flow in the left ventricle. The model successfully reproduced the physiological opening and closing of the mitral valve leaflets and simulated normal and prolapsed valve states. Additionally, it simulated blood flow patterns after the insertion of a prosthetic valve with and without obstruction of left ventricular outflow tract flow.
Background: Fluid structure interaction simulations h hold promise in studying normal and abnormal cardiac function, including the effect of fluid dynamics on mitral valve (MV) leaflet motion. The goal of this study was to develop a 3D fluid structure interaction computational model to simulate bileaflet MV when interacting with blood motion in left ventricle (LV).Methods: The model consists of ideal geometric-shaped MV leaflets and the LV, with MV dimensions based on human anatomical measurements. An experimentally-based hyperelastic isotropic material was used to model the mechanical behaviour of the MV leaflets, with chordae tendineae and papillary muscle tips also incorporated. LV myocardial tissue was prescribed using a transverse isotropic hyperelastic formulation. Incompressible Navier-Stokes fluid formulations were used to govern the blood motion, and the Arbitrary Lagrangian Eulerian (ALE) method was employed to determine the mesh deformation of the fluid and solid domains due to trans-valvular pressure on MV boundaries and the resulting leaflet movement.Results: The LV-MV generic model was able to reproduce physiological MV leaflet opening and closing profiles resulting from the time-varying atrial and ventricular pressures, as well as simulating normal and prolapsed MV states. Additionally, the model was able to simulate blood flow patterns after insertion of a prosthetic MV with and without left ventricular outflow tract flow obstruction. In the MV-LV normal model, the regurgitant blood flow fraction was 10.1 %, with no abnormality in cardiac function according to the mitral regurgitation severity grades reported by the American Society of Echocardiography.Conclusion: Our simulation approach provides insights into intraventricular fluid dynamics in a contracting LV with normal and prolapsed MV function, as well as aiding in the understanding of possible complications after transcatheter MV implantation prior to clinical trials.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据