4.6 Article

Performance of digital morphology analyzer CellaVision DC-1

期刊

CLINICAL CHEMISTRY AND LABORATORY MEDICINE
卷 61, 期 1, 页码 133-141

出版社

WALTER DE GRUYTER GMBH
DOI: 10.1515/cclm-2022-0829

关键词

comparison; DC-1; digital morphology analyzer; performance; turnaround time

向作者/读者索取更多资源

This study comprehensively evaluated the performance of CellaVision DC-1, including its turnaround time. DC-1 showed excellent precision and qualitative performance, making it suitable for small to medium-volume laboratories. However, in some cell classes, DC-1 may add unnecessary workload for cell verification.
Objectives CellaVision DC-1 (DC-1, Sysmex, Kobe, Japan) is a newly launched digital morphology analyzer that was developed mainly for small to medium-volume laboratories. We evaluated the precision, qualitative performance, comparison of cell counts between DC-1 and manual counting, and turnaround time (TAT) of DC-1. Methods Using five peripheral blood smear (PBS) slides spanning normal white blood cell (WBC) range, precision and qualitative performance of DC-1 were evaluated according to the Clinical and Laboratory Standards Institute (CLSI) EP15-A3, EP15-Ed3-IG1, and EP12-A2 guidelines. Cell counts of DC-1 and manual counting were compared according to the CLSI EP 09C-ED3 guidelines, and TAT of DC-1 was also compared with TAT of manual counting. Results DC-1 showed excellent precision (%CV, 0.0-3.5%), high specificity (98.9-100.0%), and high negative predictive value (98.4-100.0%) in 18 cell classes (12 WBC classes and six non-WBC classes). However, DC-1 showed 0% of positive predictive value in seven cell classes (metamyelocytes, myelocytes, promyelocytes, blasts, plasma cells, nucleated red blood cells, and unidentified). The largest absolute mean differences (%) of DC-1 vs. manual counting was 2.74. Total TAT (min:s) was comparable between DC-1 (8:55) and manual counting (8:55). Conclusions This is the first study that comprehensively evaluated the performance of DC-1 including its TAT. DC-1 has a reliable performance that can be used in small to medium-volume laboratories for assisting PBS review. However, DC-1 may make unnecessary workload for cell verification in some cell classes.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据